Clarity - Getting Started Connecting Autosamplers
49
utilize the OUT
n
R digital output cable. All
timings are governed by the chromatograph;
Clarity only performs one analysis for each
start signal received. All synchronization only
includes external start up of data acquisition
in Clarity using the IN
n
starting cable.
Fig. 20. Wiring of an autosampler in Passive Sequence
A Passive Sequence has to be used e.g. in the
sets with Headspace autosamplers (without AS
Control module).
Caution! It is not recommended to use the Passive
Sequence together with the Control modules.
Note: Detailed information on using Passive Sequence
can be found in the Clarity User Guide.
6.1.4 AS – Active Sequence + AS Control + A/D card
When using the optional AS Control (p/n A26)
module, all communication is performed through
a separate data cable (usually a serial cable
connected to a COM port).
Caution! Refer to the corresponding Clarity Control
manual for wiring specific to your control
module.
Following scheme corresponds to the set with
directly controlled autosampler without using the
digital acquisition, when data are acquired using
an A/D converter (INT7, U-PAD).
In such a case do NOT CONNECT
the digital
output cable OUT
n
R.
The starting cable IN
n
MUST BE CONNECTED.