Clarity - Getting Started Connecting Autosamplers
46
A
UTOSAMPLER
INJECTION
READY
CHROMATOGRAPH
READY
START
OUTPUT
START
INPUT
EXT.
START/STOP
OUTPUT
INT 7
(U-PAD, INT 5)
PC
Wiring of the autosampler,
that will be started by
the contact
at the READY input.
closing
Starting cable -
IN
n
Digital output -
OUT
n
R
Fig. 17. Wiring of the autosampler – variant A
The injection will start only after both serially
connected contacts (Clarity and the GC) are
closed.
After an injection, the autosampler will close
the INJECTION contact and thus the command
to start the temperature gradient program will
be given. At the same time, the chromatograph
will close the START contact and thus the
command to start the Clarity station
acquisition will be given.
If the chromatograph does not have a START
OUTPUT contact then the starting cable IN
n
must be connected directly to the INJECTION
output on the autosampler (this way, in fact,
parallel to the START INPUT contact of the
chromatograph).
To have the contact on the INT7 board (INT5 or
U-PAD) opened in the initial state, it is
necessary to set the Output Initial item in the
Digital Output Control dialog (accessible from the
Clarity main window) to HIGH.
Note: The start output mapping of the Clarity to
individual digital outputs of the A/D converter can
be set in the bottom-right corner of the
System
Configuration
dialog (see Fig. 2 on pg. 15). A
detailed description of this can be found in the
Reference Guide.