Clarity - Getting Started First “Analysis”
41
5.6.1 Opening a calibration file
When a calibration file was specified in the
template method used for acquiring the
chromatogram, you can open it from the
Chromatogram window.
Use the View button in the
Chromatogram -
Results
window.
Use the
File - Open Standard (yellow icon)
command to open an integrated chromatogram
representing a calibration standard.
Click the Add All
(blue line) icon to
automatically transfer available data on each
peak to the calibration table.
In the displayed calibration table, modify
compound names in the first column
.
Note: Table fields can be edited after clicking on them
with the mouse.
Fill in the known quantity for each compound
in the Amount column
under the number of
the current calibrated level.
Note: If you have performed all steps as described, you
should see a straight-line single-point calibration
curve after clicking on the tab of any individual
compound.
Finally, save the created calibration curves
using the File - Save command.
5.6.2 Link the calibration file with chromatogram or template
method
Comments
The calibration can be attached (applied) to a
chromatogram by selecting its file in the
Chromatogram – Results window (in Calibration
File (Peak Table) field.
Zobrazit stránku 40